Leica CM1950 manual Connection to Power Supply System, Existing power system

Page 24

6.Installation

6.4Connection to Power Supply System

After transporting, wait at least 4 hours before turning the instrument on. This waiting period is necessary to allow the compressor oil, which may have been displaced during transport, to re- turn into its original position. Furthermore, any condensation that has formed during this time due to temperature fluctuations must be allowed to dry completely.

Failure to comply with this can cause severe damage to the instrument!

During the start-up of the compressor the nominal voltage must not drop below the values specified in the "Technical Data".

Please note that the compressor requires a start-up current between 25 and 35 A.

Therefore, the electric circuit at the installation site must be inspected by an electrical engineer to ensure that it meets the requirements for a smooth operation of the instrument. Failure to comply with the above will cause severe damage to the instrument!

Check mains voltage and mains frequency to comply with the specifica- tion on the type plate.

Do not connect any other consumers to this electric circuit.

Never connect the instrument to a power socket that does not have a protective conductor terminal.

Only for instruments sold in Japan

 

Selecting the frequency

 

• After unpacking the instrument and setting it up at its intended location,

 

use the lever (1) to select the frequency corresponding to the conditions of

1

the existing power system.

Power plug

Fig. 19

24

Operating Manual V1.1 – 05/2007

Image 24
Contents Leica CM1950 Page Page Table of contents 11.1 12.112.2 12.3Symbols Used in This Manual and Their Meaning Important InformationQualification of Personnel Designated UseSafety and Design General Safety InstructionsUnpacking and Installation Opening the packaging Ramp assembly Transport to the Installation LocationRemoving the packaging Safety Features Microtome knivesKnife guard Securing/Locking the HandwheelCleaning, Disinfection Turning the Instrument Back On Maintenance Handling Specimens DefrostingRemoval of the Microtome Technical Data Technical Data Refrigeration system CM1950, 50 Hz CM1950, 60 HzStandard Delivery Set of filters 5 piecesHandwheel, motorized General Overview Control Panel Fields and Cryostat Chamber Installation Site RequirementsRight side Left sideTransport to the Site Assembling the Handwheel Locking/Unlocking the Handwheel Centering the Handwheel OptionalEmergency-stop function active, or Connection to Power Supply System Existing power systemOnly for instruments sold in Japan Selecting the frequency Installation Installing the Storage Systems Optional For reasons Shelf, Movable OptionalInserting the Section Waste Tray Section waste tray with the a Cutout a facing the userInstalling the Heat Conductor Block, Stationary Optional Installation Mark for suction nozzle Instrument Controls Control Panel Fields on the CM1950 Control Panel Field12 o’clock Position 6 o’clock Retraction 0μm20.0μm 60.0μmCoarse feed functions Motion continues as long as the button is heldBrief press of the button retracts the specimen by 20μm Control panel 3 Motorized sectioning optional Sectioning modesDisinfection Duration 30min Duration 180 min Leica Cryofect is not available in all countriesWorking with the Instrument Turning the Instrument OnConfiguring the Parameters Working with the Instrument Working with the Instrument Manual defrosting of the specimen cooling Entering the trimming thickness Working with the Precooled Cryostat 8.4.1 Preparatory Work Trimming with Extraction Anti-roll Guide Installed Handwheel position 12 3 o’clock, valve open all the way Cutting with Extraction Anti-roll Guide installed Shut off VAC LED in VAC button goes outSpeed 10μm 40% 60% 80% 100% 20μm 30μm 40μmSpeed Extract. value 1μm 5 μm 40% 60% 80% 100% 10μm Working with the Instrument Troubleshooting Problems During WorkProblem Causes Remedies Cryochamber Sections tear Alternately Longer sufficient Temperature Selection Chart in minus C Tissue type Chamber temperatureOptional Accessories Ordering InformationCryofect, VPE Leica Cryofect 4x 350ml When using low-profile blades, the ruler With anti-roll guideLever 6 for the lateral shift Lever 7 for clamping the bladeBlade holder CE with finger rest for NB + BB Inserting the blade into knife holder CE Will result Cleaning the knife holder CE Daily cleaning Unscrew the bolts 35 of the pressure plateNot Visible Optional Accessories Anti-roll guide system Optional Accessories Specimen discs in various dimensions Optional Accessories Maintenance and Cleaning General MaintenanceChanging Fuses Replacement of the Lamp/UVC Lamp Installation of the new lamp This equipment has been cleaned and decontaminated Decontamination Certificate masterNB Microtome knives must be in boxes Yes No indicated belowBU Return Authorization Number Anne De Greef-Safft President Biosystems Division EC Declaration of ConformityWarranty and Service Peoples Republic of China