Section 6 | Calibration |
OVERVIEW
FULLY AUTOMATIC CALIBRATION
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page
During a calibration, two calibration gases with known O2 concentrations and one calibration gas with a known COe concentration are applied to the OCX 8800. Slope and constant values are calculated to determine if the OCX 8800 is correctly measuring net concentrations of O2 and combustibles in the industrial process.
Before calibrating the OCX 8800, verify that the calibration gas parameters are correct by setting the test gas values used when calibrating the unit. Refer to Section 3, Configuration and Startup.
There are three calibration methods available to the OCX 8800; automatic,
If the OCX 8800 is equipped with calibration solenoids, the unit can be programmed to automatically calibrate without any operator action. Refer to the following paragraphs for using the LOI or HART/AMS to set up the OCX 8800 for fully automatic calibration.
Autocalibration Setup using the LOI
Use the following procedure to set up the OCX 8800 for automatic calibration. If necessary, use the LOI menu tree in Figure
NOTE
Automatic calibration is only available on units equipped with calibration solenoids.
1.From the operating display use the
2.From the SYSTEM first column submenu, use the
3.From the Calib Setup second column submenu, use the