Install Eductor

OCX 8800

Figure 7-13. Installation of Eductor and COe Sensor

8

11 9

10

12

14

7

If installed, the O2 cell and heater strut assembly (9, Figure 7-3) must be removed from sensor housing (8), before you install the eductor.

1.Apply pipe thread sealant (Loctite #567) to the external pipe threads of eductor (6, Figure 7-13). Do not apply sealant to the first turn of the pipe threads.

4

3

5

1

2

15

6

13

16

1. Insulator

2. Dilution Air Tube

3. COe Extractive Tube

4. Eductor Air Tube

5. COe Sensor Assembly

6. Eductor

7. Sensor Housing

8. COe Thermocouple

9. Heater Insulator

10. COe Band Heater

11. Sensor Holder

12. Eductor Elbow

13. Terminal Block Mounting

14. Eductor Holder

15. Tube Fitting

16. CJC Sensor