Instruction Manual

IM-106-880, Rev 1.0 January 2007

Figure 4-4. LOI Menu Tree (Sheet 2 of 4)

OCX 8800

(CONTINUED FROM

SHEET 1)

CALIBRATION Cal Control

(CONTINUED ON

SHEET 3)

Cal

Constants

Cal

Status

Cal

Verify

Abort Cal

Start Cal

O2

Start Cal

COe

Start Cal

Both

Current

Calib

Prev

Calib

Falied

Calib

Calib Result

Calib Step

Calib Time

Next O2 Cal

Next COe Cal

Flow High Gas

Flow Low Gas

Flow COe Gas

Purge

Status

Apply Lo O2 Gas

 

 

 

 

Hit E when ready

 

 

 

 

Flow O2 Lo

_____s

 

 

 

______%

_____mV

 

 

 

Read O2 Lo

_____s

 

 

 

______%

_____mV

 

 

 

Apply Hi O2 Gas

 

 

 

 

Hit E when ready

 

 

 

 

Flow O2 Hi

_____s

 

 

 

______%

_____mV

 

 

 

Read O2 Hi

_____s

 

 

 

______%

_____mV

 

 

 

Hit E when ready

 

 

 

 

Purge

_____s

 

 

 

 

 

 

 

 

 

 

 

Apply Lo COe Gas

 

 

 

Hit E when ready

 

 

 

 

Flow COe Lo

_____s

 

 

 

______%

_____mV

 

 

 

Read COe Lo _____s

 

 

 

______%

_____mV

 

 

 

Apply Hi COe Gas

 

 

 

 

Hit E when ready

 

 

 

 

Flow COe Hi

_____s

 

 

 

______%

_____mV

 

 

 

Read COe Hi _____s

 

 

 

______%

_____mV

 

 

 

Hit E when ready

 

 

 

 

Purge

_____s

 

 

 

 

 

 

 

 

 

 

 

Apply Lo O2 Gas

 

Hit E when ready

Hit E when ready

 

Purge

_____s

Flow O2 Lo

_____s

Flow COe Hi

_____s

______%

_____mV

______%

_____mV

Read O2 Lo

_____s

Read COe Hi _____s

______%

_____mV

______%

_____mV

Apply Hi O2 Gas

 

Hit E when ready

Hit E when ready

 

Purge

_____s

Flow O2 Hi

_____s

 

 

 

______%

_____mV

 

 

 

Read O2 Hi

_____s

 

 

 

______%

_____mV

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O2 Slope

 

_____mV/D

 

 

O2 Constant

 

_____mV

 

 

O2 Sensor R

 

_____ohm

 

 

COe Slope

 

_____ppm/mV

 

 

COe Constant

 

_____mV

 

 

 

 

 

 

 

Pre O2 Slope

 

_____mV/D

 

 

Pre O2 Const

 

_____mV

 

 

Pre O2 Sensor R

_____ohm

 

 

Pre COe Slope

_____ppm/mV

 

 

Pre COe Const

_____mV

 

 

 

 

 

 

 

Bad O2 Slope

 

_____mV/D

 

 

Bad O2 Const

 

_____mV

 

 

Bad COe Slope

_____ppm/mV

 

 

Bad COe Const

_____mV

 

 

NOTE: “Hit E when ready” is displayed during semi-automatic calibration only (when Calib Setup value Use Solenoids = n).

37390015

4-5

Page 45
Image 45
Emerson OCX 8800 instruction manual LOI Menu Tree Sheet 2