OCX 8800
Instruction Manual
4.Scroll down to the last item Use Solenoids. If the unit is equipped with calibration solenoids and timed automatic calibration is desired, select Yes.
5.Use the
6.Use the
7.Use the
8.Use the
Autocalibration Setup using HART
NOTE
Automatic calibration is only available on units equipped with calibration solenoids.
Use the following procedure to specify a time interval (in hours) at which the OCX 8800 will automatically calibrate.
1.From the DEVICE SETUP screen, select DETAILED SETUP.
2.From the DETAILED SETUP screen, select O2 CALIB PARAMS or
COE CALIB PARAMS.
3.If the unit is equipped with calibration solenoids and timed automatic calibrations are desired, select Solenoids, then select Yes. Selct No to disable the calibration solenoids.