Emerson OCX 8800 Figure B-3. SPA Jumper Dip Switch Settings, Configuration/Calibration

Models: OCX 8800

1 168
Download 168 pages 13.3 Kb
Page 159
Image 159

Instruction Manual

IM-106-880, Rev 1.0 January 2007

OCX 8800

Figure B-3. SPA Jumper

and Dip Switch Settings

NOTICE

REMOVE THIS PANEL FOR SERVICE ACCESS.

STATIC SENSITIVE. USE CAUTION WHEN CHANGING JUMPERS

Access Cover

SPA - Bottom View

Figure B-4. SPA Setup for Calibration

Password

Jumper Pins

 

Password

Password

 

37020022

 

 

 

Source Current Dip Switches

ON

OFF

Failsafe Dip Switches

 

(shown in correct position)

 

 

(shown in correct position)

 

2.

Verify that the Password Jumper is set to the OFF position. If the jumper

 

is in the ON position, reposition the jumper.

3.

Check the position of the Failsafe Dip Switches. Position the dip

OCX 8800

Optional

>250

<950

IN

Fluke Model 87

Multimeter

or equivalent

mA

IN

 

switches as shown in Figure B-3.

4.

Check the position of the Source Current Dip Switches. Position the dip

 

switches as shown in Figure B-3.

5.

Reinstall the SPA access cover.

Configuration/Calibration

Prior to operation, the SPA operating parameters must be configured via a menu-driven setup procedure. At the end of the configuration procedure, the SPA analog output signal is calibrated to insure valid communications.

1.See Figure B-4. Connect a calibrated ammeter (Fluke Model 87 or equivalent, accurate to ±0.025%) to the SPA analog output terminals. Observe polarity.

2.Connect a 90 to 260 VAC or 22 to 300 VDC power source to the SPA power terminals. When connecting an AC power source, use the AC and ACC (AC Common) terminals. For a DC source, use the AC and Ground terminals.

3.If desired, you can connect the 4 to 20 mA O2 signal wires from the OCX 8800 analog output terminal block to the SPA Input terminals. (The OCX must be operational to transmit the O2 signal. Observe polarity.)

 

GND

37390024

AC or DC

AC

 

 

 

 

 

 

ACC

 

 

 

Power Input

 

 

 

 

 

 

 

 

 

 

 

B-3

Page 159
Image 159
Emerson OCX 8800 instruction manual Figure B-3. SPA Jumper Dip Switch Settings, Configuration/Calibration