FAULT ISOLATION
Faults in the OCX 8800 Transmitter are indicated by messages displayed on the HART communicator or LOI. Fault indications that can appear are listed in Table
If a fault is indicated on the HART communicator or LOI, locate the fault indication in Table
Table
O2 Sensor R High (Oxygen sensor resistance high, > 5000 Ohms) O2 Sensor Open (Oxygen sensor disconnected)
Probable Cause
Recommended Corrective Action
Loose or open O2 cell circuit connection | Check O2 cell circuit wires for breaks or loose connections. Repair lead wire break or |
| loose connections. |
O2 cell degraded or failed | Check O2 cell impedance by reading the O2 Cell Imped value via the LOI (see |
| Figure |
| impedance is zero, replace O2 cell with cell replacement kit. If cell impedance is less |
| than 5000 ohms, check for cell housing ground fault. Repair ground fault. If cell |
| impedance is greater than 5000 ohms and no ground fault is indicated, replace O2 cell |
| with cell replacement kit. |
|
|
Ref Current Err (RTD excitation current error)
Probable Cause
Recommended Corrective Action
Loose or open lead or circuit wire connection for COe sensor or CJC sensor current loop
COe sensor grounded
Check all COe and CJC sensor current loop wiring per Figure
Check resistance of COe sensor leads to ground per Figure
COe sensor failed | Check resistance of both COe sensor elements per Figure |
| resistance of sensor element is not between 100 and 250 Ohms. |
CJC sensor grounded | Check resistance of CJC sensor to ground per Figure |
| resistance is less than 10M Ohms. |
CJC sensor failed | Check resistance of CJC sensor per Figure |
| sensor is not between 100 and 150 Ohms. |
O2 Temp Hi (Oxygen sensor heater temperature high, > 750ºC)
Probable Cause
Recommended Corrective Action
High noise in OCX power supply | Check power supply for line noise or voltage fluctuations. Install power line filter kit |
| (PN 6A00171G01) or high quality line filter for input power. |
COe Temp Hi (Combustion sensor heater temperature high, 310ºC)
Probable Cause
Recommended Corrective Action
High noise in OCX power supply | Check power supply for line noise or voltage fluctuations. Install power line filter kit |
| (PN 6A00171G01) or high quality line filter for input power. |
|
|
| Table continued on next page |