Instruction Manual
IM-106-880, Rev 1.0
January 2007
6-3
OCX 8800
4. Select O2 CalIntrvl (O2 calibration interval) and enter the desired time
in hours between automatic O2 calibrations. Select COE Callintrvl and
enter the desired time between automatic COe calibrations. To disable
automatic calibration for O2 and COe, enter 0 for both CalIntrvl
parameters.
5. If desired, the O2NxtCalTm and the COeNxtCalTm (next calibration
time) parameters can be changed to synchronize a calibration at a
specific day or time.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.
To return to a preceding menu, press the left arrow key.
6. From the O2 CALIB PARAMS screen, select CalIntrvl.
7. At the prompt, input a time interval (in hours) at which an automatic O2
calibration will occur and press ENTER.
8. From the DETAILED SETUP screen, select COE CALIB PARAMS.
9. From the COE CALIB PARAMS menu, select CalIntrvl.
10. At the prompt, input a time interval (in hours) at which an automatic COe
calibration will occur and press ENTER.
When setting automatic calibration times, CalIntrvl and NxtCalTm should be set so that O2
and COe are NOT calibrated simultaneously.