Instruction Manual

IM-106-880, Rev 1.0 January 2007

OCX 8800

Section 2

Installation

Mechanical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-2

Electrical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-7

Pneumatic Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-10

Initial Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-16

Before installing this equipment, read the "Safety instructions for the wiring and installation of this apparatus" in Appendix A: Safety Data. Failure to follow the safety instructions could result in serious injury or death.

To maintain explosion-proof protection of the OCX 8800 in hazardous areas, all cable entry devices and blanking elements for unused apertures must be certified flameproof, suitable for the conditions of use and properly installed.

To maintain explosion-proof protection of the OCX88C in hazardous areas, the sensor housing must not be mounted to any surface or flange that exceeds 200ºC (392ºF).

To maintain explosion-proof protection of the OCX88C in hazardous areas, the sample entering the sensor housing must not exceed 200ºC (392ºF).