Instruction Manual
OCX 8800
SPECIFICATIONS
Hazardous Area OCX
Specifications |
|
Net O2 Range | |
Combustibles | |
Accuracy |
|
Oxygen | ± 0.75% of reading or 0.05% O2 (whichever is greater) |
Combustibles | ± 2% range |
|
|
System Response to |
|
Test Gas |
|
Oxygen | 10 sec T90 |
Combustibles | 25 sec T90 |
Temperature Limits |
|
Process | 32° to 2600°F (0° to 1427°C) |
Sensors Housing | |
Electronics Housing | |
| |
| electronics inside instrument housing, as measured by a HART |
| communicator or AMS software |
Local Operator | |
Interface | [At temperatures above 158°F (70°C) inside instrument housing, |
| the infrared keypad will cease to function, but the OCX 8800 will |
| continue to operate properly.] |
Nominal and Approximate |
|
Shipping Weights |
|
18 in. (457 mm) | 54 lbs (20 kg) |
probe package |
|
3 ft (0.91 m) probe | 55 lbs (20.5 kg) |
package |
|
6 ft (1.83 m) probe | 57 lbs (21 kg) |
package |
|
9 ft (2.74 m) probe package
Housings Mounting
Materials
Probes
Enclosures
Calibration
Calibration Gas Mixtures
Recommended
(Ref. test gas bottles
kit #1A99119G04)
Calibration Gas Flow
Reference Air
Eductor Air
Dilution Air
Blowback Air (optional)
59 lbs (22 kg)
Flange
316L stainless steel - 1300°F (704°C) Inconel 600 - 1832°F (1000°C) Ceramic - 2600°F (1427°C)
0.4% O2, Balance N2
8% O2, Balance N2
1000 ppm CO, Balance Air
7 scfh (3.3 l/m)
2 scfh (1 l/m), clean, dry
5 scfh (2.5 l/m), clean, dry,
0.1scfh (0.5 l/m), clean, dry,
Clean, dry,
Table continued on next page