SB TC Open (Sample block heater thermocouple open)

OCX 8800

Probable Cause

Recommended Corrective Action

Sample block thermocouple or thermocouple circuit open

Check sample block thermocouple and circuit wires for breaks or loose connections per Figure 8-3 and Figure 7-4. Repair breaks or loose connections or replace failed thermocouple. Perform Reset procedure in Section 3, Configuration and Startup to continue operation.

O2 TC Shorted (Oxygen sensor heater thermocouple shorted)

Probable Cause

Recommended Corrective Action

O2 thermocouple or thermocouple circuit shorted

Check O2 thermocouple and circuit wires for short circuit condition per Figure 8-3 and

 

Figure 7-4. Repair shorted wiring or replace failed thermocouple. Perform Reset

 

procedure in Section 3, Configuration and Startup to continue operation.

 

Perform Reset procedure in Section 3, Configuration and Startup to continue operation.

Slow heatup during cold start

 

If alarm persists, refer to the O2 Htr Failure alarm procedure.

 

 

COe TC Shorted (Combustibles sensor heater thermocouple shorted)

Probable Cause

Recommended Corrective Action

Combustibles sensor thermocouple shorted

Check thermocouple resistance (lead to ground) of reference and active thermocouples

 

per Figure 8-3 and Figure 7-5. If either thermocouple is open or shorted to ground,

 

replace combustibles sensor.

 

Perform Reset procedure in Section 3, Configuration and Startup to continue operation.

Slow heatup during cold start

 

If alarm persists, refer to the COe Htr Failure alarm procedure.

 

 

SB TC Shorted (Sample block heater thermocouple shorted)

Probable Cause

Recommended Corrective Action

Sample block thermocouple or thermocouple circuit shorted

Check sample block thermocouple and circuit wires for short circuit condition per Figure 8-3 and Figure 7-4. Repair shorted wiring or replace failed thermocouple. Perform Reset procedure in Section 3, Configuration and Startup to continue operation.

Slow heatup during cold start

Perform Reset procedure in Section 3, Configuration and Startup to continue operation.

 

If alarm persists, refer to the SB Htr Failure alarm procedure.

 

 

O2 TC Reversed (Oxygen sensor heater thermocouple reversed)

Probable Cause

Recommended Corrective Action

O2 thermocouple wires reversed

Check O2 thermocouple wiring per Figure 8-3 and Figure 7-4. Check the wiring at the

 

sensor and inside the electronics housing. Correct reversed-wires fault. Perform Reset

 

procedure in Section 3, Configuration and Startup to continue operation.

COe TC Reversed (Combustibles sensor block heater thermocouple reversed)

Probable Cause

Recommended Corrective Action

Combustibles thermocouple wires reversed

Check combustibles thermocouple wiring per Figure 8-3 and Figure 7-5. Check the

 

wiring at the sensor and inside the electronics housing. Correct reversed-wires fault.

 

Perform Reset procedure in Section 3, Configuration and Startup to continue operation.

SB TC Reversed (Sample block heater thermocouple reversed)

Probable Cause

Recommended Corrective Action

Sample block thermocouple wires reversed

Check sample block thermocouple wiring per Figure 8-3 and Figure 7-4. Check the

 

wiring at the sensor and inside the electronics housing. Correct reversed-wires fault.

 

Perform Reset procedure in Section 3, Configuration and Startup to continue operation.

 

 

 