Emerson OCX 8800 instruction manual Manual COe Calibration using Hart

Models: OCX 8800

1 168
Download 168 pages 13.3 Kb
Page 67
Image 67

Instruction Manual

IM-106-880, Rev 1.0 January 2007

OCX 8800

Manual COe Calibration using HART

To perform a manual COe calibration using the HART communicator or AMS, use the following procedure. If necessary, refer to Section 5, Using HART Communications, for the HART menu tree.

NOTE

To select a menu item, either use the up and down arrow keys to scroll to the menu item and press the right arrow key or use the number keypad to select the menu item number.

To return to a preceding menu, press the left arrow key.

1.From the DIAG/SERVICE menu, select CALIBRATION. From the CALIBRATION menu, select PERFORM CAL.

2.From the PERFORM CAL menu, select CAL METHODS. Select COe CALIBRATION to start COe calibration.

3.In the first COe Calibration screen, a Loop should be removed from automatic control warning appears. Remove the OCX 8800 from any automatic control loops to avoid a potentially dangerous operating condition and press OK.

4.The main Calibration screen should look like the following. Press OK to continue.

OCX: TAG NAME

STATUS: Idle

TIME REMAIN: 0s

COe: 0.20 ppm, 0.00 mV OK/NEXT to Select ABORT/CANCEL to Exit

5.From the SELECT ACTION screen, select START/NEXT CALSTEP to continue calibration, select ABORT CAL to abort calibration or EXIT CAL to exit calibration. Select one item from the list and press ENTER.

OCX: TAG NAME SELECT ACTION

1.START/NEXT CALSTEP

2.ABORT CAL

3.EXIT CAL

6.The unit samples reference air as the COe Low Gas. The Calibration Status should automatically change to ReadCOLow for a period of time. During this period, if an attempt is made to go to the next calibration step by pressing OK and selecting Start/Next Cal Step, you will be prompted with Operator step command is not accepted at this time. The Next Cal Step command is not accepted at this time.

7.When ready, Calibration Status will stop at the AppCOeHi. Switch on the COe High Gas. Verify the COe concentration measured matches the COe HIGH GAS parameter in the Setup. Press OK when ready.

8.Select Start/Next Cal Step to start applying the COe High Gas. The time to apply the test gas is specified by the Gas Time.

6-7

Page 67
Image 67
Emerson OCX 8800 instruction manual Manual COe Calibration using Hart