Disassemble COe Sensor Assembly
1.Carefully remove screws (1, Figure
2.If damaged, use the following procedure to remove thermocouple adaptor (6) from sensor holder (4):
a.Use a propane torch to heat the thermocouple adaptor to 450°F (232°C), minimum.
b.While heating, use a
c.Use MEK or methylene chloride solvent to clean thread sealant residue from the internal pipe threads in the sensor holder. Refer to applicable MSDS sheet for solvent handling precautions.
Figure 7-11. COe Sensor,
Exploded View
13
2
7
6 | 9 |
8
5
4 | 1. | Screw |
| 2. | Lockwasher |
| 3. | COe Sensor |
| 4. | Sensor Holder |
| 5. | Gasket |
| 6. | Thermocouple Adapter |
| 7. | |
| 8. | Plug |
| 9. | Stainless Steel Balls |
37390030