OCX 8800
Instruction Manual
SB Temp Hi (Sample block heater temperature high, > 190ºC)
Probable Cause
Recommended Corrective Action
High noise in OCX power supply | Check power supply for line noise or voltage fluctuations. Install power line filter kit |
| (PN 6A00171G01) or high quality line filter for input power. |
O2 Temp Very Hi (Oxygen sensor heater over maximum temperature, > 820ºC)
O2 Htr Rmp Rate (Oxygen sensor heater over maximum temperature ramp rate)
Probable Cause
Recommended Corrective Action
Incorrect O2 heater wiring | Check O2 heater wiring per Figure |
| and inside the electronics housing. Correct wiring fault. Perform Reset procedure in |
| Section 3, Configuration and Startup to continue operation. |
| Check O2 thermocouple wiring per Figure |
Incorrect O2 thermocouple wiring | |
| thermocouple and inside the electronics housing. Correct wiring fault. Perform Reset |
| procedure in Section 3, Configuration and Startup to continue operation. |
Electronics package failure | Replace electronics package. |
COe Temp Very Hi (Combustion sensor heater over maximum temperature, > 400ºC) COe Htr Rmp Rate (Combustibles sensor heater over maximum temperature ramp rate)
Probable Cause
Recommended Corrective Action
Incorrect COe heater wiring | Check COe heater wiring per Figure |
| and inside the electronics housing. Correct wiring fault. Perform Reset procedure in |
| Section 3, Configuration and Startup to continue operation. |
Incorrect COe thermocouple wiring | Check COe thermocouple wiring per Figure |
| thermocouple and inside the electronics housing. Correct wiring fault. Perform Reset |
| procedure in Section 3, Configuration and Startup to continue operation. |
Electronics package failure | Replace electronics package. |
SB Temp Very Hi (Sample block heater over maximum temperature, > 260ºC)
SB Htr Rmp Rate (Sample block heater over maximum temperature ramp rate)
Probable Cause
Recommended Corrective Action
Incorrect sample block heater wiring | Check sample block heater wiring per Figure |
| the heater and inside the electronics housing. Correct wiring fault. Perform Reset |
| procedure in Section 3, Configuration and Startup to continue operation. |
Incorrect sample block thermocouple wiring | Check sample block thermocouple wiring per Figure |
| wiring at the thermocouple and inside the electronics housing. Correct wiring fault. |
| Perform Reset procedure in Section 3, Configuration and Startup to continue operation. |
Electronics package failure | Replace electronics package. |
O2 TC Open (Oxygen sensor heater thermocouple open)
Probable Cause
Recommended Corrective Action
O2 thermocouple or thermocouple circuit open | Check O2 thermocouple and circuit wires for breaks or loose connections per Figure |
| and Figure |
| Perform Reset procedure in Section 3, Configuration and Startup to continue operation. |
COe TC Open (Combustibles sensor heater thermocouple open)
Probable Cause
Recommended Corrective Action
Combustibles sensor thermocouple open | Check thermocouple resistance (lead to ground) of reference and active thermocouples |
| per Figure |
| replace combustibles sensor. |
|
|
| Table continued on next page |