Instruction Manual
OCX 8800
7.Press the
8.Press the
9.Press the
10.When the purge period expires, the LOI display reverts to the normal operation display. If the calibration failed, the display will indicate an alarm condition.
11.Press the
12.Press the
13.Press the
14.Turn off the CO test gas and press the
15.When the purge period expires, the LOI display reverts to the normal operation display. If the calibration failed, the display will indicate an alarm condition.
Manual O2 Calibration using HART
To perform a manual O2 calibration using the HART communicator or AMS, use the following procedure. If necessary, refer to Section 5, Using HART Communications, for the HART menu tree.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the menu item and press the right arrow key or use the number keypad to select the menu item number.
To return to a preceding menu, press the left arrow key.
1.From the DIAG/SERVICE menu, select CALIBRATION. From the CALIBRATION menu, select PERFORM CAL.
2.From the PERFORM CAL menu, select CAL METHODS. Select O2 CALIBRATION to start O2 calibration.
3.In the first O2 Calibration screen, a Loop should be removed from automatic control warning appears. Remove the OCX 8800 from any automatic control loops to avoid a potentially dangerous operating condition and press OK.
4.The main Calibration screen should look like the following. Press OK to continue.
OCX: TAG NAME
STATUS: Idle
TIME REMAIN: 0s
O2: 0.4 %, 85.95mV OK/NEXT to Select ABORT/CANCEL to Exit