Install O2 Cell and Heater Strut Assembly
1.Rub a small amount of
2.Apply
Stripped threads on the O2 cell and heater strut assembly can allow gas leakage. Gas leakage can affect the O2 measurements and calibration. Avoid
3.Install O2 cell and heater strut assembly (9) in sensor housing (8). Snug up, but do not
4.Reconnect the lead wires from O2 cell, heater, and thermocouple to the sensor housing terminal blocks. Refer to Figure
5.Install reference air tube (7, Figure
Install Terminals Insulator and Cover
1.Install insulator (6, Figure
2.Position opposite side of insulator (6) and secure with related marking plate (5).
3.If removed, install cover gasket (4). Screw cover (3) onto sensor housing (8). Tighten cover firmly.
4.To comply with
5.Align locking clip (2) with gap between cover ribs.
6.Loosen screw (1) and slide locking clip (2) fully into gap between cover ribs. Retighten screw (1).
Sensor Housing Leak Test
1.Install 1/4 NPT cap on dilution air inlet fitting. Install a 1/4 NPT cap on sample tube (2, Figure
2.If not in place, install exhaust tube (3, Figure
3.Connect a calibrated manometer to the CAL GAS inlet port.
4.Connect and apply clean instrument air at 35 psig (241 kPa gage) to the instrument air inlet fitting.
5.Observe the manometer reading. The reading should be from 10 to 13 inches, Water Column. Locate and correct leaks if the reading is less than 10 inches WC.