Instruction Manual

IM-106-880, Rev 1.0 January 2007

OCX 8800

REPAIR SENSOR HOUSING

Use the following procedures to remove damaged components from the OCX 8800 sensor housing and to install new replacement parts. Disassemble the unit only as needed to replace damaged components. Use the assembly procedures that apply to install replacement parts and reassemble the unit.

Sensor Housing Disassembly

Remove Cover and Terminals Insulator

1.Loosen screw (1, Figure 7-3) and slide locking clip (2) away from cover. Retighten screw (1).

2.With two hands or strap wrench, turn cover (3) counterclockwise to loosen. Unthread and remove cover.

3.Inspect cover o-ring (4) for wear or damage. Replace cover o-ring if damaged.

4.Unsnap terminal marking plates (5) and remove terminal insulator (6).

7-5

Page 75
Image 75
Emerson OCX 8800 instruction manual Repair Sensor Housing, Sensor Housing Disassembly, Remove Cover and Terminals Insulator