6.Installation

6.5.7 Inserting/Changing the Bacteria Filter

Fig. 26

6.5.8 Assembling the Filter Bag

1

2

A

L3a

3

Fig. 27

E

Fig. 28

The holder for the bacterial filter (optional) is visible in the front of the instrument.

To insert the filter, hold it with one hand, press on the right of the socket, then guide the filter into the tube from the left.

To change the filter, follow the opposite proce- dure: press the filter to the right, then pull it to the left and out of the tube.

The filter must be changed approx. every 3 months (we recommend writing the date on the filter using a marker).

The filter must be disposed of accord- ing to valid laboratory specifications.

Set the mark (A) of the extraction opening (1) and pull it out. Plug the filter (3) into the ex- traction hose connecting piece (3a) until there is an audible click.

Now push the connected parts back into the opening in the cryostat chamber (filter first) and set it to the "closed" mark (2, Fig. 27).

When not using the extraction, tightly close the opening for the extraction hose using the silicone stopper (E) in- cluded in the scope of delivery.

Reasons:

1.To prevent section waste from falling into the opening.

2.To prevent cold from escaping from the chamber.

3.To prevent moisture from penetrating into the chamber.

28

Operating Manual V1.1 – 05/2007

Image 28
Contents Leica CM1950 Page Page Table of contents 11.1 12.112.2 12.3Symbols Used in This Manual and Their Meaning Important InformationQualification of Personnel Designated UseGeneral Safety Instructions Safety and DesignUnpacking and Installation Opening the packaging Transport to the Installation Location Ramp assemblyRemoving the packaging Safety Features Microtome knivesKnife guard Securing/Locking the HandwheelCleaning, Disinfection Turning the Instrument Back On Handling Specimens Defrosting MaintenanceRemoval of the Microtome Technical Data Technical Data Refrigeration system CM1950, 50 Hz CM1950, 60 HzStandard Delivery Set of filters 5 piecesHandwheel, motorized General Overview Control Panel Fields and Cryostat Chamber Installation Site RequirementsRight side Left sideTransport to the Site Assembling the Handwheel Centering the Handwheel Optional Locking/Unlocking the HandwheelEmergency-stop function active, or Existing power system Connection to Power Supply SystemOnly for instruments sold in Japan Selecting the frequency Installation Installing the Storage Systems Optional For reasons Shelf, Movable OptionalInserting the Section Waste Tray Section waste tray with the a Cutout a facing the userInstalling the Heat Conductor Block, Stationary Optional Installation Mark for suction nozzle Instrument Controls Control Panel Fields on the CM1950 Control Panel Field12 o’clock Position 6 o’clock Retraction 0μm20.0μm 60.0μmMotion continues as long as the button is held Coarse feed functionsBrief press of the button retracts the specimen by 20μm Control panel 3 Motorized sectioning optional Sectioning modesDisinfection Duration 30min Duration 180 min Leica Cryofect is not available in all countriesWorking with the Instrument Turning the Instrument OnConfiguring the Parameters Working with the Instrument Working with the Instrument Manual defrosting of the specimen cooling Entering the trimming thickness Working with the Precooled Cryostat 8.4.1 Preparatory Work Trimming with Extraction Anti-roll Guide Installed Handwheel position 12 3 o’clock, valve open all the way Cutting with Extraction Anti-roll Guide installed Shut off VAC LED in VAC button goes out10μm 40% 60% 80% 100% 20μm 30μm 40μm SpeedSpeed Extract. value 1μm 5 μm 40% 60% 80% 100% 10μm Working with the Instrument Problems During Work TroubleshootingProblem Causes Remedies Cryochamber Sections tear Alternately Longer sufficient Temperature Selection Chart in minus C Tissue type Chamber temperatureOptional Accessories Ordering InformationCryofect, VPE Leica Cryofect 4x 350ml When using low-profile blades, the ruler With anti-roll guideLever 6 for the lateral shift Lever 7 for clamping the bladeBlade holder CE with finger rest for NB + BB Inserting the blade into knife holder CE Will result Cleaning the knife holder CE Daily cleaning Unscrew the bolts 35 of the pressure plateNot Visible Optional Accessories Anti-roll guide system Optional Accessories Specimen discs in various dimensions Optional Accessories Maintenance and Cleaning General MaintenanceChanging Fuses Replacement of the Lamp/UVC Lamp Installation of the new lamp This equipment has been cleaned and decontaminated Decontamination Certificate masterNB Microtome knives must be in boxes Yes No indicated belowBU Return Authorization Number Anne De Greef-Safft President Biosystems Division EC Declaration of ConformityWarranty and Service Peoples Republic of China