2.Safety and Design
2.5Cleaning, Disinfection – Turning the Instrument Back On
It is not necessary to remove the microtome for disinfection.
•The instrument has been designed for UV disinfection! Spray disinfec- tion with Leica Cryofect is also possible, thanks to the special insulation of the microtome. (Cryofect is not available in all countries!)
Remove section waste after EVERY sectioning operation and BE- FORE changing specimens. Remove the section waste using a pa- per towel soaked in Cryofect or an
•When disinfecting the instrument, take appropriate protective measures (gloves, mask, protective clothing, etc.).
•When using detergents and disinfectants please comply with the safety precautions of the disinfectant manufacturer.
•The integrated glass
•Dispose of waste liquid according to the waste disposal regulations!
•Do not use external heaters for drying the cryochamber. This can cause damage to the cooling system!
•Do not turn the instrument on before the cryochamber is completely dry! Frost formation!
•All components removed from the cryostat must be carefully dried be- fore returning them to the cryochamber!
•The front panel and the slit cover of the microtome must be completely dry before turning on the instrument!
For more detailed information about disinfection, visit the Leica
Microsystems website at
12 | Operating Manual V1.1 – 05/2007 |