Models: 965AMS

1 204
Download 204 pages 1.32 Kb
Page 62
Image 62
2.Press the blue  key to enter the RFL function.

Measurement Functions

3M™ Dynatel™ Advanced Modular System 965AMS

RFL>Separate-Pair, Single-Section Operation

2.Press the blue Manual backgroundManual background key to enter the RFL function.

Press Select until the Fault section is highlighted. Use the up and down arrow keys to select the type of fault

that you have. The options include:

Tip Cross, Ring Cross, Wet

Pulp, T-G, R-G, Short

This example will show a Ring-Ground fault.

Press Select to move to the Pair section.

3.Use the up and down arrow keys to select Separate Pair.

Press Select to move to the Multiple-Single section.

62