If you discover incomplete characters, numbers, or symbols

!

on the display, put your insulin pump into STOP mode and

CAUTION

perform a system check by simply removing the battery for a

 

few seconds. When you reinsert the battery the system check

 

will occur. If the incomplete characters, numbers, or symbols

 

remain, your insulin pump may need to be returned since you

 

may not receive the correct information for operating your

 

pump. Put your insulin pump into STOP mode and contact

 

Pump Support at 1-800-688-4578 immediately.

Note:

To make the display easier to read, you can flip the display orientation by 180° and adjust the display contrast lighter or darker. See the sections “Display Orientation” (page 120) and “Display Contrast” (page 135) for more information.

Backlight

The backlight helps you to see the display in poorly lighted areas. You can turn on the backlight from the RUN screen, from the STOP screen, or when you are browsing the menus. Press the a key to turn on the backlight. The backlight turns itself on when an alert or an error message appears. The backlight turns itself off after 30 seconds if no keys are pressed.

Hint:

Start programming with the a also turn on the backlight.

key in areas of poor lighting as this will

Chapter 1: Understanding Your New System

10

ACCU-CHEK Spirit Insulin Pump

Page 11
Image 11
Accu-Chek insulin pump manual Backlight, Hint, Pump Support at 1-800-688-4578 immediately