Technical Standards on Electromagnetic Immunity

Guidance and manufacturer’s declaration – electromagnetic immunity

The ACCU-CHEK Spirit insulin pump is intended for use in the electromagnetic environment specified below. The customer or the user of the ACCU-CHEK Spirit insulin pump should assure that it is used in such an environment.

 

IEC 60601

Compliance

Electromagnetic

Immunity Test

Test Level

Level

Environment – Guidance

 

 

 

 

Electrostatic

± 6 kV contact

± 8 kV contact

Avoid any contact with

discharge (ESD)

± 8 kV air

± 15 kV air

synthetic materials. A

IEC 61000-4-2

 

(IEC 60601-2-24) relative humidity of 10% will

 

 

 

allow ESD to be more likely.

 

 

 

 

Power frequency

3 A/m 400 A/m

400 A/m

Power frequency magnetic

(50/60Hz)

 

 

fields should be at levels

magnetic field

 

 

characteristic of a typical

IEC 61000-4-8

 

(IEC 60601-2-24)

location in a typical

 

 

 

commercial or hospital

 

 

 

environment.

 

 

 

 

Conducted RF

3 Vrms

Not applicable

 

IEC 61000-4-8

150 kHz to

 

 

 

80 MHz

 

 

 

 

 

 

Appendixes

171

ACCU-CHEK Spirit Insulin Pump

Page 172