Programming a Standard Bolus

There are two ways to program a Standard Bolus:

“Quick” Standard Bolus

“Scroll” Standard Bolus

Using a and s on your

Menu-guided using d and f

insulin pump, with bolus increments

with a and s for

determined by the user

programming of the amount

The bolus amount per delivery is limited to 25 units using U100 insulin.

After you program a Standard Bolus there is a delay of 5 seconds before the bolus delivery starts. This delay lets you cancel the bolus before delivery starts by pressing the a or s key. Alert A8: BOLUS CANCELLED will occur. Press f twice to confirm and turn off the alert. See the section “Alert A8: BOLUS CANCELLED” (page 153) for more information.

Programming a “Quick” Standard Bolus

The following process involves only the a and s keys. You can start with either key. The recommended steps below are those described with the first key appearing without parenthesis, such as a. For the alternative step press the key in the parenthesis, such as (s).

Note:

Because your pump beeps and vibrates to guide you safely through the programming, you may program the “Quick” STANDARD Bolus using one finger. After the bolus is programmed, you can count the beeps and vibrations of your pump to verify the correct programming.

The bolus increment for the “Quick” Standard Bolus is initially set to 0.5 units.

Chapter 4: Boluses and TBRs

77

ACCU-CHEK Spirit Insulin Pump

Page 78
Image 78
Accu-Chek insulin pump manual Programming a Standard Bolus, There are two ways to program a Standard Bolus