IEC 60601

Compliance

Electromagnetic

Immunity Test

Test Level

Level

Environment – Guidance

 

 

 

 

Radiated RF

3 V/m

10 V/m

Portable and mobile RF

 

 

 

communications equipment

IEC 61000-4-3

80 MHz to

80 MHz

should be used no closer to

 

2.5 GHz

2.5 GHz

any part of the ACCU-CHEK

 

 

 

Spirit insulin pump, includ-

 

 

(IEC 60601-2-24) ing cables, than the recom-

 

 

 

mended separation distance

 

 

 

calculated from the equation

 

 

 

applicable to the frequency

 

 

 

of the transmitter.

 

 

 

Recommended separation

 

 

 

distance:

 

 

 

 

s

80 MHz

 

 

 

d = 0.4 .EP

 

 

 

s

to 800 MHz

 

 

 

800 MHz

 

 

 

d = 0.7 .EP

to 2.5 GHz

where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey (a), should be less than the compliance level in each frequency range (b).

Interface may occur in the vicinity of equipment marked with the following symbol: