1.Press f twice to turn off and to confirm the error.

2.Remove or disconnect the infusion set from your infusion site.

3.Prime the infusion set.

4.Reconnect the infusion set (tubing) at your infusion site.

5.Put your pump in RUN mode.

See the sections “Changing the Cartridge and Infusion Set” (page 73) and “Priming the Infusion Set” (page 53) for more information.

Error E12: DATA INTERRUPTED

If data transfer between compatible software and your

!

insulin pump is disrupted in any way, the configuration

CAUTION

may be incomplete and an error E12: DATA INTERRUPTED

 

occurs. The data transfer must be completed successfully

 

before you can put your insulin pump into RUN mode.

Data transfer between your pump and a PC was interrupted before it was completed.

1.Press f twice to turn off and to confirm the error.

2.Restart the data transfer.

Before you can put your pump in RUN mode, the data transfer must be successful.

Chapter 7: Alerts and Errors