RUN

3 sec.

bolus

screen

 

amount

3 sec.

bolus

if O unit

5 sec. delay

delivery

 

confirmation

 

 

 

 

 

 

of amount

 

 

 

5 sec. delay

A8: bolus cancelled

Follow the steps below:

1.From the RUN screen press and hold s (a) until you hear a beep sequence and feel a vibration.

This activates the “Quick” STANDARD

BOLUS function.

Hint:

Start programming with the a key in areas of poor lighting, as this will also turn on the backlight.

2.Press a (s) repeatedly until the bolus amount is reached.

Each press of the a (s) key adds one bolus increment to the bolus amount. Your pump beeps and vibrates simultaneously once for each programmed bolus increment.

Five seconds after the last press of a (s), your pump confirms the total bolus amount with one beep and vibration for each bolus increment programmed.

The Standard Bolus symbol ( ) blinks for 5 seconds (the bolus delivery start delay).

After this bolus delivery start delay, the pump beeps three times and then begins to deliver the total bolus amount programmed. The countdown of the remaining bolus amount appears on the display.

Chapter 4: Boluses and TBRs

78

ACCU-CHEK Spirit Insulin Pump

Page 79
Image 79
Accu-Chek insulin pump manual Run