Error E3: AUTOMATIC OFF

Automatic off is a safety feature that stops insulin delivery and triggers an error E3: AUTOMATIC OFF if no keys are pressed within a programmed time period in RUN mode.

1.Press f twice to turn off and to confirm the error.

2.Put your pump in RUN mode.

Error E4: OCCLUSION

If an error E4: OCCLUSION occurs, immediately check your

!

blood glucose level, and check again in about 2 hours,

WARNING

because insulin delivery has been interrupted. If your blood

 

glucose level is high, take appropriate actions according to

 

your doctor’s or healthcare team’s instructions.

Insulin (a maximum of 3.5 units of U100 insulin) was not delivered. This may be due to:

a blocked infusion set,

a reused cartridge, or

a dirty or damaged piston rod.

Note:

The display of this error depends on several factors including your current hourly basal rate and boluses. For example, a blocked infusion set with a basal rate of 2.0 U/h (U100 insulin and no additional boluses) would trigger an error E4: OCCLUSION in approximately three hours or less, depending on the type and severity of the blockage.

Chapter 7: Alerts and Errors

156

ACCU-CHEK Spirit Insulin Pump

Page 157
Image 157
Accu-Chek insulin pump Error E3 Automatic OFF, Error E4 Occlusion, Your doctor’s or healthcare team’s instructions, 156