Alert A4: ALARM CLOCK

An alarm clock reminder was programmed into the pump.

1.Press f twice to turn off and to confirm the alert.

2.If necessary, put your pump in RUN mode.

Note:

If you have set the alarm occurrence to EVERY DAY, the same alert occurs every 24 hours. Turn the alarm clock function OFF if you do not want the alarm to go off again at that same time every day.

See the section “Alarm Clock” (page 127) for more information.

Alert A5: PUMP TIMER

Your pump’s operating time will expire soon. It is recommended that you replace your pump after a period of four years of running time. After the alert A5: PUMP TIMER occurs, the ACCU-CHEK Spirit insulin pump will display the symbol to remind you to check the timer.

1.Press f twice to turn off and to confirm the alert.

2.Put your pump in RUN mode.

3.Go to the INFORMATION screen on your pump.

Chapter 7: Alerts and Errors

151

ACCU-CHEK Spirit Insulin Pump

Page 152
Image 152
Accu-Chek insulin pump manual Alert A4 Alarm Clock, See the section Alarm Clock page 127 for more information, 151