Note:

If a Temporary Basal Rate is active at the same time, it is also cancelled when your pump is put into STOP mode. An alert A8: BOLUS CANCELLED and an alert A6: TBR CANCELLED occur.

Press f twice to confirm and turn off the first alert. The second alert appears on the display. Press f twice to confirm and turn off the second alert. Both alerts are recorded in the alarm history.

See the sections “Alert A6: TBR CANCELLED” (page 152) and “Alert A8: BOLUS CANCELLED” (page 153) for more information.

Make sure that the cancellation was intended and then program a new Extended Bolus (and/or Temporary Basal Rate), if needed.

Chapter 4: Boluses and TBRs

87

ACCU-CHEK Spirit Insulin Pump

Page 88
Image 88
Accu-Chek insulin pump manual Boluses and TBRs