Alert A7: TBR OVER (Temporary Basal Rate Over)

A Temporary Basal Rate has ended. The original basal rate (100%) will resume automatically once the alert has been confirmed.

1.Press f twice to turn off and confirm the alert.

2.Decide whether a further Temporary Basal Rate change is needed. If necessary, program a new Temporary Basal Rate.

Alert A8: BOLUS CANCELLED

A bolus was cancelled during the start delay or after delivery began.

1.Press f twice to turn off and to confirm the alert.

2.If necessary, put your pump in RUN mode.

3.Make sure that the cancellation was intended and then program a new bolus if needed.

Note:

The actual bolus amount that was delivered before the cancellation can be reviewed in the bolus history. See the section “Review the Bolus History” (page 107) for more information.

Chapter 7: Alerts and Errors

153

ACCU-CHEK Spirit Insulin Pump