Error E10: CARTRIDGE ERROR

The CHANGE THE CARTRIDGE function was not correctly carried out.

1.Press f twice to turn off and to confirm the error.

2.Remove or disconnect the infusion set from your infusion site.

3.Start the CHANGE THE CARTRIDGE menu.

4.Prime the infusion set.

5.Reconnect the infusion set (tubing) at your infusion site.

6.Put your pump into RUN mode.

7.If an error E10: CARTRIDGE ERROR occurs during the winding of the piston rod, insert a new battery into the pump and repeat steps 1-6.

Error E11: SET NOT PRIMED

Never prime an infusion set that is connected to your body.

!

You risk uncontrolled insulin delivery into your body. With

WARNING

detachable infusion sets, make sure you have disconnected

 

the tubing from your site prior to priming. Always follow

 

the instructions provided with your infusion set.

The cartridge was changed but the infusion set was not primed. The error will display when you try to put your pump in RUN mode.