Follow the instructions below:

Remove the protective cap from the cartridge. Do not touch the cartridge tip.

b

-Llockue connecto

r

a

Push the adapter all the way onto the cartridge tip to the stop (a).

Hold the adapter and twist the infusion set luer-lock connection by hand, clockwise into the adapter (b). Carefully prepare a new ACCU-CHEK infusion set for use, priming and insertion, following its instructions for use.

 

Do not over-tighten.

To avoid leakage, properly tighten the infusion set to the

!

adapter. Turn the infusion set luer-lock connector until it

CAUTION

stops. Do not turn it any further and do not use any auxiliary

 

tools as this may crack the infusion set luer-lock connector

 

and result in leakage.

Note:

Your pump cannot detect infusion set leakage. You must check all parts of the infusion set at least every three hours during the day and before you go to sleep. If you detect any loss of insulin and all parts are correctly tightened, immediately replace the leaking component. Because insulin delivery has been interrupted, check your blood glucose level and take appropriate actions according to your doctor’s or healthcare team’s instructions.

Chapter 2: Getting Started

48

ACCU-CHEK Spirit Insulin Pump

Page 49
Image 49