8.Prime the infusion set following all the instructions and warnings found in the section “Priming the Infusion Set” (page 53).

9.Insert a new infusion set following the instructions found in the section “Preparing the Infusion Site and Inserting the Infusion Set” (page 56).

10.Remove the old infusion headset and dispose of it properly.

11.Place the pump in RUN mode. See the section “Starting Insulin Delivery” (page 65).

Note:

If you use a soft cannula-type infusion set, you must proceed with a bolus to fill the air space in the cannula after removing the introducer needle according to the instructions for the infusion set you are using. Failure to do so could result in missed insulin dosing. See the section “Programming a Standard Bolus” (page 77) for more information.

Inspect your infusion site at least once or twice each day for irritation or infection. Signs of infection may include, but are not limited to, pain, lumps, redness, heat, or discharge.

If you see redness or swelling, change the infusion headset and the infusion site immediately and contact your doctor or healthcare team.

Change the infusion headset according to the instructions for the infusion set you are using and your doctor or healthcare team’s recommendations.

Chapter 3: Using Your Pump

72

ACCU-CHEK Spirit Insulin Pump

Page 73
Image 73
Accu-Chek insulin pump manual Using Your Pump