Problem

Recommended Actions

Occlusion in system

Disconnect and re-prime your infusion

 

set. Check your blood glucose. If the

 

occlusion remains, prime the pump

 

without the cartridge inserted. See the

 

section “Error E4: OCCLUSION” (page 156).

Failure to complete prime

Re-prime the infusion set. Always prime

 

until a bubble-free insulin flow emerges

 

from the tip of the needle. Make sure the

 

tubing is free of air.

Poor absorption of insulin

Choose infusion sites that do not have

 

scar tissue, bruises, or tissue build-up.

 

Do not use the same infusion set or

 

infusion site longer than your doctor or

 

healthcare team recommends.

Infusion site becomes sore,

Change your infusion headset and infusion

red, or swollen.

site immediately. Use proper insertion

 

technique. Check blood glucose level.

 

Follow the planned infusion site rotation

 

pattern and advice given by your doctor or

 

healthcare team.