Accuracy of Flow Rate for Plastic Cartridges

Trumpet curve plotted from data after the end of the stabilization period

The trumpet curve shows the accuracy of the delivery rate in relation to the observation period.

The measurements were made at a medium basal rate of 1.0 U/h according to IEC 60601-2-24:1998 with a Disetronic Classic infusion set PC 16/110 at room temperature.

The average deviation of the delivered amount (overall mean percentage flow error) for U100 insulin is ± 5%.

Appendixes

173

ACCU-CHEK Spirit Insulin Pump

Page 174
Image 174
Accu-Chek insulin pump manual Accuracy of Flow Rate for Plastic Cartridges, 173