Lock Basal Rate Profiles

Your pump lets you change your hourly basal rates as needed. You may lock your basal rate profiles to provide extra protection against accidentally changing a basal rate during normal operation. When this function is activated, your hourly basal rates cannot be changed.

Follow the steps below:

1.Press d to move to the SETUP MENU ADVANCED screen.

Press f to select.

2.Press d to move to the BASAL RATE LOCK screen.

Press aor sto turn the basal rate lock ON or OFF.

3.Press f to save and exit.

Note:

If the basal rate lock is set to ON ( ), the programming of the basal rate profiles 1, 2, 3, 4, and 5 is blocked. To check if the basal rate lock is ON, press d to move to the PROGRAM BASAL RATE screen and press f to select. The symbol in the BASAL RATE TOTAL screen indicates the lock of the basal rate programming. (The symbol appears in the BASAL RATE TOTAL screen only.) Press d to view the hourly basal rates. The a and s functions are blocked.