Accu-Chek insulin pump Therapy-Related Issues, Troubleshooting Table Problem Recommended Actions

Models: insulin pump

1 200
Download 200 pages 16.16 Kb
Page 145
Image 145

Therapy-Related Issues

Many factors, such as alcohol consumption, medications other than insulin, ineffective or expired insulin, decreased activity, illness, and stress can affect blood glucose levels. Contact your doctor or healthcare team for advice on how to manage these or any other therapy related issues.

Troubleshooting Table

Problem

Recommended Actions

Individual characters, numbers, or

Contact Pump Support at

symbols on the display are shown

1-800-688-4578.

incompletely or not at all.

 

Your insulin pump does not beep

Check that the beep volume and alarm

or vibrate.

signals are turned on. Check and

 

change the battery. If your insulin pump

 

still does not beep or vibrate, contact

 

Pump Support at 1-800-688-4578.

Self-test incomplete (after an error

Remove the battery, wait a few

E7: ELECTRONIC ERROR).

seconds, and insert it again. If the

 

self-test remains incomplete, contact

 

Pump Support at 1-800-688-4578.

Incorrect insulin delivery due to

The correct hourly basal rate cannot be

incorrect time setting.

delivered if the time is not set correctly.

 

Check and reset the time if it is wrong.

 

Do the same check when you change

 

the battery.

Incorrect insulin delivery due to

Always review changes made in basal

incorrect basal rate programming.

rate programming. Confirm that you are

 

using your correct basal rate profile.

Chapter 6: Care and Maintenance

144

ACCU-CHEK Spirit Insulin Pump

Page 145
Image 145
Accu-Chek insulin pump manual Therapy-Related Issues, Troubleshooting Table Problem Recommended Actions, 144