Flow (delivery rate)

0.2 U/sec during infusion set

 

priming and bolus when using

 

U100 insulin.

 

 

Infusion sets

ACCU-CHEK infusion sets with

 

luer-lock connector.

 

 

 

 

IPX 8

Protected against the effects of

 

temporary immersion in water

 

under standardized conditions

 

[up to 60 minutes and 8 feet

 

 

(2.5 meters)].

 

 

 

 

 

 

Maximum time for an error

Plastic cartridges*

 

 

E4: OCCLUSION

• at med. basal rate 1.0 U/h using

 

U100 insulin

5h

 

• at min. basal rate 0.1 U/h using

 

U100 insulin

50h

Maximum volume* before error

U100 insulin

3.5 U

Maximum pressure

400kPa (4.0 bar)

1.0 U

Maximum quantity delivered at

U100 insulin

a single fault condition

 

 

 

 

 

Power supply

One 1.5 volt AA Alkaline battery

 

or one rechargeable Nickel Metal

 

Hydride (NiMH) AA battery. Alkaline

 

batteries should have a minimum

 

capacity of 2500 mAh and NiMH

 

AA rechargeable batteries a

 

 

minimum capacity of 1500 mAh.

 

Use only a battery charger

 

 

 

officially recommended by the

 

battery manufacturer.

 

 

Pump casing

Shock, scratch, pharmaceutical

 

resistant plastic. All edges are

 

rounded.

 

 

*Determined using measurement methods according to IEC 60601-2-24-1998.