Cancelling an Extended Bolus

during programming (the bolus amount or bolus duration blinks):

You have three options for cancelling an Extended Bolus during programming:

If you do not press any key for 20 seconds, your pump returns to the RUN screen (timeout).

Exit to the EXTENDED BOLUS menu by pressing d+asimultaneously.

Set the bolus amount to 0.0 units and press f.

No bolus is delivered.

during bolus delivery:

If the Extended Bolus delivery has begun, it can be cancelled by putting your pump into STOP mode. This cancels the bolus delivery and an alert A8: BOLUS CANCELLED occurs. Press f twice to confirm and turn off the alert. See the section “Alert A8: BOLUS CANCELLED” (page 153) for more information.

Bolus delivery is interrupted.

The bolus amount delivered up to the cancellation can be reviewed in the BOLUS HISTORY screen. See the section “Review the Bolus History” (page 107) for more information.

When your insulin pump is in STOP mode, it does not deliver

!

any insulin. Put the pump in RUN mode to continue the

CAUTION

insulin delivery.

Chapter 4: Boluses and TBRs

86

ACCU-CHEK Spirit Insulin Pump

Page 87