If the error E6: MECHANICAL ERROR occurs again in spite of these actions, contact your doctor or healthcare team for an alternate therapy plan. Contact Pump Support at 1-800-688-4578 for more assistance.

Error E7: ELECTRONIC ERROR

An electronic error was discovered during a routine automatic system check and the insulin delivery has stopped. You cannot turn off and confirm an error E7: ELECTRONIC ERROR by pressing f. You must remove the battery instead.

1.Remove or disconnect the infusion set from your infusion site.

2.Insert a new battery into the pump.

3.Prime the infusion set.

4.Reconnect the infusion set (tubing) at your infusion site.

5.Put your pump in RUN mode.

If an error E7: ELECTRONIC ERROR occurs during programming (i.e., basal rate, time and date) check your settings.

If the error E7: ELECTRONIC ERROR occurs again, contact Pump Support at 1-800-688-4578 for assistance. Contact your doctor or healthcare team if you need an alternate therapy plan.