Inserting the Cartridge

Do not perform the CHANGE THE CARTRIDGE function if the

!

cartridge compartment is not completely dry. During the

CAUTION

rewinding of the piston rod, insulin or water might enter your

 

insulin pump and could lead to malfunction. If necessary,

 

use a soft cloth to dry the compartment.

STOP

 

change the

remove the 3 sec.

screen

cartridge

cartridge

 

 

 

returning

 

 

 

piston rod

 

 

 

cartridge volume

 

 

 

315 units

 

 

 

start-up

 

 

 

procedure

 

 

prime the

 

 

 

infusion set

 

Have the following materials ready:

Your insulin pump

A filled ACCU-CHEK 3.15 ml plastic cartridge with a connected ACCU-CHEK Spirit insulin pump adapter

A new ACCU-CHEK infusion set

Chapter 2: Getting Started

49

ACCU-CHEK Spirit Insulin Pump

Page 50
Image 50
Accu-Chek insulin pump manual Inserting the Cartridge, Use a soft cloth to dry the compartment