Error E2: BATTERY DEPLETED

The battery is out of power. Replace it immediately.

After inserting a new battery, always check that the time

!

and date of your insulin pump are set correctly. Incorrect

WARNING

programming of the time and date may cause incorrect

 

insulin delivery. If you, your doctor, or healthcare team

 

store and analyze your therapy data electronically and the

 

time and date of the devices used are not set identically,

 

the gathered data might not be meaningful.

1.Press f twice to turn off and to confirm the error.

2.Replace the battery.

3.Put your pump in RUN mode.

Note:

Your pump stores the settings you saved before the battery power ran out. If your pump is without power for more than one hour, check the time and date.

See the section “Accessories and Disposables–Battery” (page 21) for more information on recommended batteries. See the section “Inserting (or changing) the Battery” (page 25) for instructions on how to change your battery.