Remove the plunger rod by turning it counter-clockwise out of the plunger.

Remove the filling needle with its protective cap from the cartridge by turning counter-clockwise. Dispose of properly.

Push the cartridge protective cap tightly onto the cartridge tip until it clicks.

The cartridge is now ready to use.

Note:

Do not pull or push the plunger rod during removal.

Air bubbles in the cartridge and infusion set may cause

!

the infusion of air instead of insulin. In this case, your body

CAUTION

does not receive the required amount of insulin and an

 

error E4: OCCLUSION may be displayed. Remove bubbles

 

while filling the cartridge and priming the infusion set and

 

without having the infusion set connected to your body.

 

Cold insulin may release air when warmed. Use only insulin

 

at room temperature when filling the cartridge and priming

 

the infusion set. Inspect the cartridge and the infusion set for

 

air bubbles at least every three hours during the day and

 

before you go to sleep. Remove any air bubbles and change

 

system components if needed.

Chapter 2: Getting Started

46

ACCU-CHEK Spirit Insulin Pump