Problem

Recommended Actions

Air bubbles in the cartridge

Check blood glucose level. In the

and/or infusion set

cartridge: Disconnect the infusion tube

 

set from the headset and re-prime your

 

infusion tube set. If the air bubbles

 

remain, replace the cartridge. Make sure

 

to fill the cartridge with insulin at room

 

temperature. In the infusion set:

 

Disconnect and prime your infusion set.

 

Reconnect and make sure all connections

 

are tight and secure.

Blood in the infusion set tubing

Change your infusion set and infusion site.

Empty cartridge

Replace the cartridge. Cartridges are for

 

single use only.

Disconnected or dislodged

Make sure all connections are tight. Inspect

infusion set

connections for leaking insulin. Check your

 

blood glucose level. Change your infusion

 

set and infusion site.

Leak in system

Check for leaking insulin at all connections

 

and on the skin. Change your infusion set,

 

infusion site, cartridge, and adapter. Check

 

your blood glucose level.

Chapter 6: Care and Maintenance

145