Agilent Technologies L4400 manual Y1152A Switch Control

Models: L4400

1 279
Download 279 pages 27.3 Kb
Page 177
Image 177

Microwave Switch/Attenuator Driver

7

16 Conductor Cable

Item

Description

Example Part Numbers

 

 

 

 

Cable Type

16 conductor ribbon cable, 0.050" pitch, 26 or

3M

3801/16 (26 AWG)

 

28 AWG stranded

3M

3365/16 (28 AWG)

 

 

 

 

Y1152A Connector

10 pin socket connector, 0.1" x 0.1" pin grid,

3M

P/N 89116-0101

 

IDC termination, center polarizing key

AMP P/N 76288-3

 

 

 

 

Switch Connector

16 pin socket connector, 0.1" x 0.1" pin grid,

3M

P/N 89116-0101

 

IDC termination, center polarizing key

AMP P/N 76288-3

 

 

 

 

Cable Wiring

Y1152A connector pin 1 to switch connector

 

 

 

pin 1

 

 

 

 

 

 

9 Conductor Cable

Item

Description

Example Part Numbers

 

 

 

 

Cable Type

9 conductor ribbon cable, 0.050" pitch, 26 or

3M

3801/09 (26 AWG)

 

28 AWG stranded

3M

3365/09 (28 AWG)

 

 

 

 

Y1150A Connector

10 pin socket connector, 0.1" x 0.1" pin grid,

3M

P/N 89110-0101

 

IDC termination, center polarizing key

AMP P/N 76288-1

 

 

 

 

Switch Connector

9 pin D-sub male, IDC termination, without

3M

P/N 8209-6000

 

threaded insert

AMP P/N 747306-4

 

 

 

 

Cable Wiring

Y1152A socket connector pin 1 to switch

 

 

 

D-sub connector pin 1

 

 

 

(Note: pin 10 of Y1152A connector not used)

 

 

 

 

 

 

Y1152A Switch Control

All channels are driven in PAIRed mode.

 

Path closed*

Path open*

 

 

 

SW1 Path1

ROUT:OPEN (@xx01)

ROUT:CLOS (@xx01)

 

 

 

SW1 Path2

ROUT:OPEN (@xx02)

ROUT:CLOS (@xx02)

 

 

 

SW1 Path3

ROUT:OPEN (@xx03)

ROUT:CLOS (@xx03)

 

 

 

SW1 Path4

ROUT:OPEN (@xx04)

ROUT:CLOS (@xx04)

 

 

 

SW1 Path5

ROUT:OPEN (@xx05)

ROUT:CLOS (@xx05)

 

 

 

SW1 Path6

ROUT:OPEN (@xx06)

ROUT:CLOS (@xx06)

 

 

 

 

State A

State B

 

 

 

SW2

ROUT:OPEN (@xx07)

ROUT:CLOS (@xx07)

 

 

 

SW3

ROUT:OPEN (@xx08)

ROUT:CLOS (@xx08)

 

 

 

* For switches connected to SW1, note the path closed is accomplished with the ROUTe:OPEN command.

L4400 User’s Guide

167

Page 177
Image 177
Agilent Technologies L4400 manual Y1152A Switch Control