Software Installation and Configuration

2

Verifying the GPIB Path

In the Agilent Connection Expert window, select and open ‘Interactive IO’. Verify communication to the instrument by sending the *IDN? command using Send & Read below the command line.

Figure 2-18. GPIB Communication Using the Default GPIB Address.

The GPIB Address String

When programming the L4400 instruments over GPIB, the instrument’s GPIB address is included in the address string. For example:

GPIB0::9::INSTR

Changing the GPIB Address

If you have only one L4400 instrument on the GPIB interface and there are no other instruments on the bus, the L4400 instrument address can remain set to 9. If you have multiple L4400 instruments or there is another GPIB instrument at address 9, then one of the addresses must be changed.

The command used to set the GPIB address on all L4400 instruments is:

SYSTem:COMMunication:GPIB:ADDRess < address >

The command can be abbreviated by including only the upper-case letters in the com- mand syntax. The Interactive IO window is used to set the GPIB address as shown in

L4400 User’s Guide

35

Page 45
Image 45
Agilent Technologies L4400 manual Gpib Address String, Verifying the Gpib Path, Changing the Gpib Address