Microwave Switch/Attenuator Driver

7

87222 Cable

Item

Description

Example Part Numbers

 

 

 

 

Cable Type

10 conductor ribbon cable, 0.050" pitch, 26 or

3M

3801/10 (26 AWG)

 

28 AWG stranded

3M

3365/10 (28 AWG)

 

 

 

 

Y1154A Connector

14 pin socket connector, 0.1" x 0.1" pin grid,

3M

P/N 89114-0101

 

IDC termination, center polarizing key

AMP P/N 76288-2

 

 

 

 

Switch Connector

10 pin socket connector, 0.1" x 0.1" pin grid,

3M

P/N 89110-0101

 

IDC termination, center polarizing key

AMP P/N 76288-1

 

 

 

 

Cable Wiring

Y1154A connector pin 1 to switch connector

 

 

 

pin 1 (Note: pins 11 - 14 of 14 pin connector not

 

 

 

used)

 

 

 

 

 

 

9 Conductor Cable

Item

Description

Example Part Numbers

 

 

 

 

Cable Type

9 conductor ribbon cable, 0.050" pitch, 26 or

3M

3801/09 (26 AWG)

 

28 AWG stranded

3M

3365/09 (28 AWG)

 

 

 

 

Y1154A Connector

10 pin socket connector, 0.1" x 0.1" pin grid,

3M

P/N 89110-0101

 

IDC termination, center polarizing key

AMP P/N 76288-1

 

 

 

 

Switch Connector

9 pin D-sub male, IDC termination, without

3M

P/N 8209-6000

 

threaded insert

AMP P/N 747306-4

 

 

 

 

Cable Wiring

Y1154A socket connector pin 1 to switch

 

 

 

D-sub connector pin 1

 

 

 

(Note: pin 10 of Y1154A connector not used)

 

 

 

 

 

 

Y1154A Switch Control

All channels are operated in PAIRed mode.

 

State A

State B

 

 

 

SW1

ROUT:OPEN (@xx01)

ROUT:CLOS (@xx01)

 

 

 

SW2

ROUT:OPEN (@xx02)

ROUT:CLOS (@xx02)

 

 

 

SW3

ROUT:OPEN (@xx03)

ROUT:CLOS (@xx03)

 

 

 

SW4

ROUT:OPEN (@xx04)

ROUT:CLOS (@xx04)

 

 

 

SW5

ROUT:OPEN (@xx05)

ROUT:CLOS (@xx05)

 

 

 

SW6

ROUT:OPEN (@xx06)

ROUT:CLOS (@xx06)

 

 

 

SW7

ROUT:OPEN (@xx07)

ROUT:CLOS (@xx07)

 

 

 

SW8

ROUT:OPEN (@xx08)

ROUT:CLOS (@xx08)

 

 

 

L4400 User’s Guide

177

Page 187
Image 187
Agilent Technologies L4400 manual Y1154A Switch Control, Cable