L4433A Dual/Quad 4x8 Reed Matrix

5

Warning -the insulation of the wiring used with the terminal block must be rated for the highest voltage that will be present on the terminal block or on the analog bus.

Although columns are numbered the same on Matrix 1 and Matrix 2, they are electrically separate from one another (e.g., Col C2).

COLUMN

When using the 34933T terminal block for 2- wire mode, access is provided to the bypass columns through the columns labeled C9 through C16. Follow this wiring convention shown in the table below for both matrices.

Terminal marked...

Connects to...

Terminal marked...

Connects to...

 

 

 

 

C9H

C1Hbypass

C13H

C5H bypass

 

 

 

 

C9L

C1L bypass

C13L

C5L bypass

 

 

 

 

C10H

C2H bypass

C14H

C6H bypass

 

 

 

 

C10L

C2L bypass

C14L

C6L bypass

 

 

 

 

C11H

C3H bypass

C15H

C7H bypass

 

 

 

 

C11L

C3L bypass

C15L

CC7L bypass

 

 

 

 

C12H

C4H bypass

C16H

C8H bypass

 

 

 

 

C12L

C4L bypass

C16L

C8L bypass

 

 

 

 

L4400 User’s Guide

123

Page 133
Image 133
Agilent Technologies manual L4400 User’s Guide 123