Agilent Technologies L4400 manual Instrument Power-On and Default LAN Configuration States

Models: L4400

1 279
Download 279 pages 27.3 Kb
Page 55
Image 55

Software Installation and Configuration

2

Instrument Power-On and Default LAN Configuration States

The L4400 series instruments covered in this user’s guide are set to their power on and preset states using any one the following commands:

*RST

SYSTem:CPON 1

SYSTem:PRESet

Table 2- 3 lists the power- on and preset states for each instrument.

Table 2-3. L4400 Instrument Power-on and Preset States

L4400 Series Instrument

Power-on Preset States

 

 

L4421A 40-Channel Armature Multiplexer

All channels open

 

 

L4433A Dual/Quad 4x8 Reed Matrix

All channels open

 

2-wire/1-wire mode: no change

 

 

L4437A 32-Channel Form A/ Form C

All channels open

General Purpose Switch

 

 

 

L4445A Microwave Switch / Attenuator Driver

Channel drives enabled = user-

 

defined defaults

 

 

L4450A 64-Bit Digital I/O w/Memory and Counter

I/O ports = Input

 

Count = 0

 

Trace memory = cleared

 

 

L4451A 4-Channel Isolated D/A Converter

DACs = 0Vdc

w/Memory

Trace wavforms = cleared

 

 

 

L4452A Multifunction with Digital I/O, D/A,

DIO Ports = Input

Totalizer

Count = 0

 

 

DACs = 0Vdc

 

 

LAN Reset (Default) Configuration

Pressing the “LAN Reset” button (recessed) on the L4400 instrument front or rear panel restores the instrument’s default LAN configuration. Table 2- 4 lists the default LAN configuration settings.

L4400 User’s Guide

45

Page 55
Image 55
Agilent Technologies L4400 manual Instrument Power-On and Default LAN Configuration States, LAN Reset Default Configuration