Microwave Switch/Attenuator Driver

7

Paired Drive With Separate Position Indicators

The simplified schematic below illustrates the connection for a dual drive switch with separate position indicators. The position indicators for this type of switch are independent relay contacts that are mechanically linked to the RF switch position.

The RF paths are not shown in the simplified diagram. The coils are driven in open collector mode. The position indicator is set so that a high level indicates an active switch. The logic level of the position indicator can be inverted using the ROUTe:CHANnel:VERify:POLarity command.

As shown, Channel 01 was pulsed to close Coil A. The corresponding position indicator also closed. Closing position indicator A opens position indicator B.

The schematic shown is similar to the Agilent N181x series of switches.

34945EXT

Y1155A

Switch

 

 

Distribution

 

 

Logic Gate

Board

 

 

 

 

 

Sense

IND 11

 

 

 

 

 

 

IND 1

 

 

Pull Down

 

 

 

Resistor

 

A

B

 

 

 

+VI

 

 

 

+VR

 

 

 

 

Coil A

Coil B

Open Collector

 

 

 

Output Drivers

DRV 1

 

 

 

 

 

 

DRV 11

 

 

L4400 User’s Guide

 

 

185

Page 195